Journal of Xidian University ›› 2016, Vol. 43 ›› Issue (4): 166-171.doi: 10.3969/j.issn.1001-2400.2016.04.029

• Article • Previous Articles     Next Articles

ECG front-end subsystem with the driven-right-leg circuit

YANG Zheng;WANG Jingmin;ZHU Zhangming;YANG Yintang   

  1. (School of Microelectronics, Xidian Univ., Xi'an  710071, China)
  • Received:2015-08-31 Online:2016-08-20 Published:2016-10-12

Abstract:

Under the condition of an un-shielding room, electrocardiogram (ECG) signal detection is a weak signal detection process with a strong common mode interference. In order to improve the performance of common mode rejection of the circuit, a driven-right-leg (DRL) circuit is often utilized. In this paper, a high performance ECG amplifier is proposed. The programmable gain amplifier (PGA) circuit and DRL circuit are applied at this analog front end. The PGA is utilized for ECG signal amplification, with the programmable gain set at 1, 2, 3, 4, 6, 8, 12. The DRL circuit is used to reduce common-mode interference.

Key words: electroence phalograph, amplifier, driven-right-leg, biopotential measurement, CMRR


Baidu
map